CUNSTCAMERA (CABINET OF CURIOSITIES) OF CLINICAL SCIENCE

Exhibit №2
Results of independent metrological testing of programs for calculation of the heart rate variability parameters

7.1.Calculation errors of spectrum power
Remember, that calculation of error is attitude of value of power the calculated program toward set in a standard sequence (1 - there is not an error)

7.1.1.Errors of calculation of general spectrum power (Total Power -TP)

The Name of file of testing sample in format: frequency (f), powery (P) P adjusted (ms2)
The coefficient of a mistake of calculation of TP
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
  fVLF_fLF_fHF_P              
1
001_007_025_2.rg
3*2
1,17
2,00
2,50
1,67
0,67
1,00
2 001_007_025_200.rg
3*200
1,05
2,01
0,54
1,49
0,97
0,99
3 001_007_025_5000.rg
3*5000
1,01
2,01
0,11
1,49
0,99
0,99
4 00032_0041_016_1000.rg
3*1000
0,99
1,83
0,49
1,42
1,01
0,97
5 0039_014_039_1000.rg
3*1000
0,97
2,03
0,49
1,16
0,89
0,99
  fVLF_P
6 001_2.rg
2
1,45
2,00
9,00
no data
1,00
1,08
7 001_200.rg
200
1,07
1,71
0,51
no data
1,08
1,00
8 001_5000.rg
5000
1,02
1,68
0,11
1,95
1,06
0,98
9 002_2.rg
2
3,75
2,50
7,65
no data
1,00
1,11
10 002_200.rg
200
1,09
2,52
0,54
2,0
1,05
1,01
11 002_5000.rg
5000
1,01
2,49
0,12
1,96
1,04
0,99
  fLF_P
12 006_2.rg
2
3,45
2,50
6,85
no data
1,00
1,12
13 006_200.rg
200
1,15
2,32
0,52
no data
1,06
1,01
14 006_5000.rg
5000
1,00
2,29
0,10
1,86
1,05
0,99
15 012_2.rg
2
4,05
2,50
6,10
no data
1,00
1,11
16 012_200.rg
200
1,21
2,54
0,60
1,70
1,02
1,01
17 012_5000.rg
5000
1,02
2,51
0,11
1,62
1,01
1,00
  fHF_P
18 017_2.rg
2
2,85
2,50
7,95
2,00
1,00
1,10
19 017_200.rg
200
1,02
2,55
0,65
1,47
0,98
1,01
20 017_5000.rg
5000
1,00
2,51
0,12
1,36
0,97
1,00
21 034_2.rg
2
2,15
3,00
5,50
0,50
0,50
1,12
22 034_200.rg
200
1,18
2,53
0,63
0,46
0,73
1,01
23 034_5000.rg
5000
0,95
2,50
0,12
0,41
0,72
1,00

Note:
- By a grey color is selected tests with the extremely low set powers which actual for the special objects of researches (look the chapter);
- no data - the program did not execute calculations with this selection

7.1.2.Errors of power calculation of spectrum are in the range of high-frequencies (HF)

The Name of file of testing sample in format: frequency (f), powery (P) P adjusted(ms2)
The coefficient of a mistake of calculation of VLF
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
  fVLF_fLF_fHF_P              
1
001_007_025_2.rg
3*2
0,35
1,50
0,28
2,00
0,50
0,93
2
001_007_025_200.rg
3*200
0,92
1,51
0,01
1,98
1,07
0,99
3
001_007_025_5000.rg
3*5000
0,96
1,50
0,01
1,94
1,07
0,99
4
00032_0041_016_1000.rg
3*1000
0,57
1,39
0,13
1,60
1,10
0,61
5
0039_014_039_1000.rg
3*1000
0,74
0,27
0,27
1,54
0,00
1,00
 
fVLF_P
6
001_2.rg
2
0,25
1,50
0,44
no data
1,00
1,05
7
001_200.rg
200
0,95
1,53
0,06
no data
1,07
0,99
8
001_5000.rg
5000
1,00
1,51
0,02
1,95
1,05
0,98
9
002_2.rg
2
0,65
2,00
0,90
no data
1,00
1,07
10
002_200.rg
200
0,97
1,76
0,45
1,99
1,04
1,00
11
002_5000.rg
5000
0,98
1,73
0,07
1,95
1,03
0,99

Note:
- By a grey color is selected tests with the extremely low set powers which actual for the special objects of researches (look the chapter);
- no data - the program did not execute calculations with this selection

7.1.3.Errors of power calculation of spectrum are in the range of low frequencies (LF)

The Name of file of testing sample in format: frequency (f), powery (P) P заданная (ms2)
The coefficient of a mistake of calculation of LF
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
  fVLF_fLF_fHF_P              
1
001_007_025_2.rg
3*2
1,45
2,00
2,21
2,00
1,00
1,04
2
001_007_025_200.rg
3*200
1,03
2,09
0,54
1,85
1,01
0,99
3
001_007_025_5000.rg
3*5000
0,96
2,10
0,10
1,84
1,01
0,99
4
00032_0041_016_1000.rg
3*1000
0,87
1,72
0,43
1,40
0,97
1,02
5
0039_014_039_1000.rg
3*1000
1,30
3,21
0,77
1,81
2,02
0,99
  fLF_P
12 006_2.rg
2
1,45
2,00
2,30
no data
1,00
1,08
13 006_200.rg
200
1,03
1,99
0,48
no data
1,06
1,00
14 006_5000.rg
5000
0,98
1,97
0,09
1,85
1,05
0,99
15 012_2.rg
2
1,60
2,00
1,25
no data
1,00
1,08
16 012_200.rg
200
1,10
2,08
0,53
1,69
1,02
1,00
17 012_5000.rg
5000
0,99
2,05
0,10
1,61
1,01
0,99

Note:
- By a grey color is selected tests with the extremely low set powers which actual for the special objects of researches (look the chapter);
- no data - the program did not execute calculations with this selection

7.1.4.Errors of power calculation of spectrum are in the range of high-frequencies (HF)

The Name of file of testing sample in format: frequency (f), powery (P) P заданная (ms2)
The coefficient of a mistake of calculation of HF
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
  fVLF_fLF_fHF_P              
1 001_007_025_2.rg
3*2
1,65
2,50
5,05
1,00
0,50
1,03
2 001_007_025_200.rg
3*200
1,18
2,44
0,65
0,65
0,85
1,00
3 001_007_025_5000.rg
3*5000
1,07
2,44
0,16
0,68
0,89
1,00
4 00032_0041_016_1000.rg
3*1000
1,01
2,37
0,50
1,27
0,96
1,01
5 0039_014_039_1000.rg
3*1000
0,87
2,60
0,52
0,13
0,66
0,99
  fHF_P
18 017_2.rg
2
1,55
2,50
5,60
2,00
1,00
1,08
19 017_200.rg
200
0,97
2,15
0,61
1,47
0,97
1,01
20 017_5000.rg
5000
0,97
2,12
0,12
1,36
0,95
1,00
21 034_2.rg
2
1,10
2,50
3,90
0,50
0,50
1,11
22 034_200.rg
200
1,08
2,37
0,60
0,46
0,73
1,01
23 034_5000.rg
5000
0,95
2,34
0,12
0,41
0,72
1,00

Note:
- By a grey color is selected tests with the extremely low set powers which actual for the special objects of researches (look the chapter);

 


Return to contents Next part