CUNSTCAMERA (CABINET OF CURIOSITIES) OF CLINICAL SCIENCE

Exhibit №2
Results of independent metrological testing of programs for calculation of the heart rate variability parameters

7.2.Non-linearity of gain-frequency descriptions
We will remind that non-linearity of AFC (amplitude-frequency characteristic) is calculated as a relation of difference of maximal and minimum values of the calculated powers of spectrum in relation to the mean value of the calculated power (0 is absolute linearness)

Numbers of test samples on which was calculated a parameter
P adjusted(ms2)
Non-linearity (%)
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
fVLF_fLF_fHF_P              
1
3*2
111,4
50,00
191
60
75,00
11,0
2
3*200
24,4
46
119
89
22,60
1,2
3
3*5000
10,7
46
135
85
17,67
1,4
4
3*1000
45,1
54
76
23
14,09
42
5
3*1000
56,9
145
101
145
225,59
1,04
fVLF_P
6 - 9
2
88,9
29
68
no data
zero
2,4
7 -10
200
1,8
13,7
152
no data
2,84
0,3
8 -11
5000
1,3
13,8
113
0,031
2,36
0,5
fLF_P
12 - 15
2
9,8
0,0
59
no data
zero
0,2
13 - 16
200
6,6
4,2
11,4
no data
3,86
0,4
14 -17
5000
1,5
4,1
6,8
14
3,84
0,3
fHF_P
18 - 21
2
34,0
0,0
36
120
66,7
2,9
19 -22
200
10,3
10,0
2,5
105
28,4
0,5
20 - 23
5000
2,3
10,1
4,1
108
28,2
0,3

Notes:
1. By a grey color is selected tests with the extremely low set powers which actual for the special objects of researches (look the chapter);
2 no data - the program did not execute calculations with this selection;
3.zero - in the results of test program reported about zero-mistake.