Results of independent metrological testing of programs for calculation of the heart rate variability parameters

7.4.Exactness of estimation of frequencies of spectrum of HRV

Exactness of calculation of frequencies of every accordion of spectral transformation is rendered by substantial influence on repetition of results at the use of the different programs. If the step of diskretization of spectral transformation is too large, then probability of casual hit of certain spectral peak, which is on the border of standard ranges rises, in one or another range. If this spectral peak of has large power, then output these presentations of parameters of spectrum of HRV are substantially distorted.
The use of left-side and right-side tests is foreseen in protocol, that in these test sequences in every spectral range powers (1000 ms2) are set on the counter-clockwise and right scopes of standard ranges.

The names of files of test sequences are in a format: frequency (f), power (P)
The calculated values of power are in a "left-side" test
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
00032_0041_016_1000.rg            
TP (ms2) 2976 5478 1465 4268 3024 2924
VLF(ms2) 566 1386 126 1599 1099 611
LF(ms2) 874 1721 434 1404 968 1024
HF(ms2) 1013 2370 498 1265 957 1006

In a left-side test from data of table some programs are calculated by the understated powers in area of very low frequencies (VLF). In particular, in the three programs: UM300, CardioLab+, MAK-3 measured VLF taking into account a bias error (for UM300 -50%, see section 7.1 Errors of calculation of spectrum power) 25% makes in relation to the set frequency, 57% and 61% accordingly. In the programs of CardioSpectrum and OrtoScience taking into account their bias error of measuring of VLF the satisfactory measuring is marked on the counter-clockwise edge of this range. Most high-quality results of this test has program CARDIO-10DX.
The graphs of spectrums are below presented in the testable programs on results treatment of this test sequence.

 

Кардиоспектр CARDIO-10DX UM300
CardioLab+ OrtoScience МАК-3

The followings results are got in a right-side test:

The names of files of test sequences are in a format: frequency (f), power (P)  

The calculated values of power are in a "right-side" test
CardioLab+ CardioSpectrum UM300 OrtoScience CARDIO-10DX МАК-3
0039_014_039_1000.rg            
TP (ms2) 2907 6083 1484 3474 2677 2977
VLF(ms2) 744 270 270 1540 3 997
LF(ms2) 1295 3211 770 1807 1016 987
HF(ms2) 866 2602 521 127 658 993

More substantial errors are exposed in this test. In the program CardioLab+ part of powers of VLF and HF went out into the test of LF. In the program CardioSpectrum VLF reckoned as LF.
In the program OrtoScience part of HF reckoned as LF. In the program CARDIO-10DX almost all of VLF reckoned as LF
.

The graphs of spectrums of implementation of this test in the different programs are presented below.

CardioSpectrum CARDIO-10DX UM300
CardioLab+ OrtoScience МАК-3

Comment:
the results of this test show existence of high probability of errors of calculation of spectral powers in different ranges at the location of measureable accordions on the scopes of spectral ranges. This methodical problem requires perfection of the programs of calculation of spectrum of HRV (for example, increase of discreteness of spectral transformation). From other side these results reflect a methodological problem which is related to the attempt of hard classification of spectral ranges.