Exhibit ¹2 Results of independent metrological testing of programs for calculation of the heart rate variability parameters |
||||||||||
3. Development of original metrology testing of HRV-programs In order to organize the testing process we need to have test sequences of RR-intervals and a device which allow to simulate these sequences like a native ECG signal. With such hardwar-program complex we get possibility to import RR-intervals signal with the set spectrum parameters directly to checked up devices and then to estimate quality of HRV-calculations of spectral parameters.3.1. Development of the program imitator TRRS Tne interface of the program-generator TRRS is rotined on a picture: The program of generation of TRRS with the set gain-frequency descriptions works as follows: three frequencies and three proper powers are set in the window of interface. Then forming of proper TRRS is then started. As a result we have a saved file "export.rg". The structure of this file is simple: duration of every RR-interval is saved in a separate line. In the present variant of the program forms 512 RR-intervals with average duration of 1000 ms.
These exemplary TRRS evidently and transparently show the work of TRRS-program of generator are a good instrument for teachingThis program is accessible by address: http://www.pallar.com.ua/files/hrv_imit.zip Every interested persons can do own experiments with this program and make sure in its high quality. Some commercial soft don`t have import procedures. At that situation entering of TRRS is possible only by using a special device which transformate of digital presentation of durations into signals in artificial ECG. 3.2. Hardwair-program complex of TRRS transformation into ECG signals Such signal allows to input without distortions TRRS into the studied devices. RR-graphics, which was got by our complex is rotined below. This is a screenshot of cardiomonitor (Utas firm) UM300: This complex was tested by our device of getting primary information and calculation of HRV-parameters. This device includes a personal computer and the program for collection and primary analysis of RR-intervals: "Monitoring system for intensive therapy and hyperbaric oxygenation", (program MAK-3) has a quality certificate on a software product 1.08.1995 MZ of Ukraine. Metrology experiment with a few TRRS showed that the registered durations of RR-intervals exactly corresponded with the digital row of TRRS and has an exactness less than the 1 msec.
|
Return to content | Next part |
---|