Return on Main Page

CUNSTCAMERA (CABINET OF CURIOSITIES) OF CLINICAL SCIENCE

Exhibit ¹2
Results of independent metrological testing of programs for calculation of the heart rate variability parameters

6.5. Primary material of self testing of the program CARDIO-10DX (Pallar LTD, Vinnitca)

Fragment of screen with HRV-conclusion and results of calculations of series N2

Results of HRV-calculations of test sequences by the program CARDIO-10DX

¹ The Name of file of testing sample in format: frequency (f), powery (P) P adjusted (ms2) P calculated (ms2)
TP VLF LF HF
  fVLF_fLF_fHF_P          
1 001_007_025_2.rg
3*2
4 1 2 1
2 001_007_025_200.rg
3*200
584 213 202 169
3 001_007_025_5000.rg
3*5000
14818 5330 5031 4457
4 00032_0041_016_1000.rg
3*1000
3024 1099 968 957
5 0039_014_039_1000.rg
3*1000
2677 3 2016 658
  fVLF_P
       
6 001_2.rg
2
2 2 0 0
7 001_200.rg
200
215 214 1 0
8 001_5000.rg
5000
5304 5269 35 0
9 002_2.rg
2
2 2 0 0
10 002_200.rg
200
209 208 1 0
11 002_5000.rg
5000
5175 5146 29 0
  fLF_P
       
12 006_2.rg
2
2 0 2 0
13 006_200.rg
200
211 0 211 0
14 006_5000.rg
5000
5227 2 5225 0
15 012_2.rg
2
2 0 2 0
16 012_200.rg
200
203 0 203 0
17 012_5000.rg
5000
5041 1 5028 12
  fHF_P
       
18 017_2.rg
2
2 0 0 2
19 017_200.rg
200
196 0 3 193
20 017_5000.rg
5000
4851 11 80 4760
21 034_2.rg
2
1 0 0 1
22 034_200.rg
200
145 0 0 145
23 034_5000.rg
5000
3595 1 11 3583

 


Return to content Next part
Hosted by uCoz