Return on Main Page

CUNSTCAMERA (CABINET OF CURIOSITIES) OF CLINICAL SCIENCE

Exhibit ¹2
Results of independent metrological testing of programs for calculation of the heart rate variability parameters

6.2. Primary material of the independent testing of the program Cardiospektrum (Solvaig)

Fragment of screen with HRV-conclusion and results of calculations of series N2

Results of HRV-calculations of test sequences by the program Cardiospektum

¹ The Name of file of testing sample in format: frequency (f), powery (P) P adjusted (ms2) P calculated (ms2)
TP VLF LF HF
  fVLF_fLF_fHF_P          
1 001_007_025_2.rg
3*2
12 3 4 5
2 001_007_025_200.rg
3*200
1207 302 418 487
3 001_007_025_5000.rg
3*5000
30208 7513 10501 12183
4 00032_0041_016_1000.rg
3*1000
5478 1386 1721 2370
5 0039_014_039_1000.rg
3*1000
6083 270 3211 2602
  fVLF_P
6 001_2.rg
2
4 3 0 0
7 001_200.rg
200
342 306 22 13
8 001_5000.rg
5000
8424 7549 552 323
9 002_2.rg
2
5 4 1 1
10 002_200.rg
200
503 351 91 61
11 002_5000.rg
5000
12425 8668 2245 1512
  fLF_P
12 006_2.rg
2
5 0 4 0
13 006_200.rg
200
464 43 398 23
14 006_5000.rg
5000
11470 1054 9838 578
15 012_2.rg
2
5 0 4 1
16 012_200.rg
200
508 9 415 83
17 012_5000.rg
5000
12542 227 10254 2060
  fHF_P
18 017_2.rg
2
5 0 1 5
19 017_200.rg
200
509 11 69 429
20 017_5000.rg
5000
12555 278 1695 10582
21 034_2.rg
2
6 0 0 5
22 034_200.rg
200
506 7 25 474
23 034_5000.rg
5000
12510 179 620 11711

 


Return to content Next part