7.3.Noise descriptions of the programs of calculation of spectrum of HRV
We will remind that non-linearity of amplitude-frequency characteristic is calculated as a relation of difference of maximal and minimum values of the calculated powers of spectrum in relation to the mean value of the calculated power (0% is absolute linearness)
№ |
The Name of file of testing sample in format: frequency (f), powery (P) |
P adjusted (ms2) |
The noise level in range of VLF (%) |
CardioLab+ |
Cardiospectrum |
UM300 |
OrtoScience |
CARDIO-10DX |
МАК-3 |
|
fLF_P |
|
|
|
|
|
|
|
12 |
006_2.rg |
2 |
35,0 |
0,00 |
35,00 |
no data |
0,00 |
0,87 |
13 |
006_200.rg |
200 |
1,8 |
21,50 |
0,45 |
no data |
0,00 |
0,62 |
14 |
006_5000.rg |
5000 |
0,8 |
21,08 |
0,70 |
1,1 |
0,04 |
0,61 |
15 |
012_2.rg |
2 |
35,0 |
0,00 |
25,00 |
no data |
0,00 |
0,37 |
16 |
012_200.rg |
200 |
1,1 |
4,50 |
0,25 |
0,5 |
0,00 |
0,16 |
17 |
012_5000.rg |
5000 |
0,2 |
4,54 |
0,01 |
0,32 |
0,02 |
0,16 |
|
fHF_P |
|
|
|
|
|
|
|
18 |
017_2.rg |
2 |
25,0 |
0,00 |
45,00 |
2,00 |
0,00 |
0,02 |
19 |
017_200.rg |
200 |
0,8 |
5,50 |
0,40 |
1,47 |
0,00 |
0,00 |
20 |
017_5000.rg |
5000 |
0,3 |
5,56 |
0,02 |
1,36 |
0,22 |
0,00 |
21 |
034_2.rg |
2 |
20,0 |
0,00 |
20,00 |
0,50 |
0,00 |
0,16 |
22 |
034_200.rg |
200 |
1,0 |
3,50 |
0,60 |
0,46 |
0,00 |
0,00 |
23 |
034_5000.rg |
5000 |
0,2 |
3,58 |
0,01 |
0,41 |
0,02 |
0,00 |
№ |
The Name of file of testing sample in format: frequency (f), powery (P) |
P adjusted (ms2) |
The noise level in range of LF (%) |
CardioLab+ |
CardioSpectrum |
UM300 |
OrtoScience |
CARDIO-10DX |
МАК-3 |
|
fVLF_P |
|
|
|
|
|
|
|
6 |
001_2.rg |
2 |
55,0 |
0,00 |
225,00 |
no data |
0,00 |
2,39 |
7 |
001_200.rg |
200 |
3,1 |
11,00 |
2,75 |
no data |
0,50 |
0,20 |
8 |
001_5000.rg |
5000 |
0,4 |
11,04 |
0,25 |
0,34 |
0,70 |
0,19 |
9 |
002_2.rg |
2 |
135,0 |
50,00 |
185,00 |
no data |
0,00 |
1,57 |
10 |
002_200.rg |
200 |
4,0 |
45,50 |
1,45 |
1,0 |
0,50 |
0,57 |
11 |
002_5000.rg |
5000 |
1,4 |
44,90 |
0,66 |
0,98 |
0,58 |
0,57 |
|
fHF_P |
|
|
|
|
|
|
|
18 |
017_2.rg |
2 |
85,0 |
50,00 |
180,00 |
0,00 |
0,00 |
1,72 |
19 |
017_200.rg |
200 |
3,8 |
34,50 |
3,40 |
0,00 |
1,50 |
0,02 |
20 |
017_5000.rg |
5000 |
1,9 |
33,90 |
0,56 |
0,02 |
1,60 |
0,01 |
21 |
034_2.rg |
2 |
75,0 |
0,00 |
110,00 |
0,00 |
0,00 |
0,46 |
22 |
034_200.rg |
200 |
4,1 |
12,50 |
2,20 |
0,00 |
0,00 |
0,01 |
23 |
034_5000.rg |
5000 |
0,5 |
12,40 |
0,08 |
0,00 |
0,22 |
0,01 |
№ |
The Name of file of testing sample in format: frequency (f), powery (P) |
P заданная (ms2) |
The noise level in range of HF (%) |
CardioLab+ |
CardioSpectrum |
UM300 |
OrtoScience |
CARDIO-10DX |
МАК-3 |
|
fVLF_P |
|
|
|
|
|
|
|
6 |
001_2.rg |
2 |
65,0 |
0,00 |
620,00 |
no data |
0,0 |
0,90 |
7 |
001_200.rg |
200 |
6,6 |
6,50 |
3,95 |
no data |
0,0 |
0,08 |
8 |
001_5000.rg |
5000 |
0,5 |
6,46 |
0,27 |
0,1 |
0,0 |
0,05 |
9 |
002_2.rg |
2 |
170,0 |
50,00 |
480,00 |
no data |
0,0 |
2,24 |
10 |
002_200.rg |
200 |
8,6 |
30,50 |
3,25 |
0,5 |
0,0 |
0,14 |
11 |
002_5000.rg |
5000 |
0,6 |
30,24 |
0,30 |
0,2 |
0,0 |
0,12 |
|
fLF_P |
|
|
|
|
|
|
|
12 |
006_2.rg |
2 |
165,0 |
0,00 |
410,00 |
no data |
0,00 |
3,02 |
13 |
006_200.rg |
200 |
9,5 |
11,50 |
4,30 |
no data |
0,00 |
0,11 |
14 |
006_5000.rg |
5000 |
0,5 |
11,56 |
0,34 |
0,24 |
0,00 |
0,08 |
15 |
012_2.rg |
2 |
210,0 |
50,00 |
450,00 |
no data |
0,00 |
2,86 |
16 |
012_200.rg |
200 |
9,1 |
41,50 |
6,10 |
0,5 |
0,00 |
0,55 |
17 |
012_5000.rg |
5000 |
1,4 |
41,20 |
0,90 |
0,82 |
0,24 |
0,52 |
Notes:
1. By a grey color is selected tests with the extremely low set powers which actual for the special objects of researches (look the chapter);
2 no data - the program did not execute calculations with this selection;
3.zero - in the results of test program reported about zero mistake.
4.Programs CardioSpectrum and CARDIO-10DX calculate spectrum power in integer format, therefore in area of extreme low powers (2ms2) the results of calculation of noise descriptions are substantially distorted
|